CVE Intelligence
Skip to main content
CRITICAL

CVE-2024-48966

CVE-2024-48966 — Life2000 service tools for test and calibration do not support user authentication

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Published Updated Sources: cvelistV5, Baxter

Triage

Is it exploited, how likely is exploitation, what does it touch, and how severe do the scoring sources call it.

Exploitation

Unreported

no source claims exploitation

EPSS

1%

chance of exploitation in 30 days

Affects

baxter

2 products listed

CVSS base

10

CRITICAL

CISA SSVC assessment

Three decision points CISA publishes for the CVEs it assesses · SSVC 2.0.3. A stakeholder decision, not a severity score.

CISA

Exploitation

None

none · proof-of-concept · active

Automatable

Yes

can an attacker script all four kill-chain steps

Technical impact

Total

partial · total control of the vulnerable component

Affected scope

The catalog records vendors and products as separate lists, not pairs, so which product belongs to which vendor is not something this page can say.

Vendors (1)

Products (2)

life2000 ventilation systemlife2000 ventilator firmware

Every base score collected

Sources score independently and disagree; each row says who scored it and under which version.

ScoreVersionSeverityExpl.ImpactSource
10CVSS 3.1CRITICALcvelistV5

Weakness & attack patterns

  • CWE-306

Detection

Read off the CVSS vector and the weakness class. Starting points, not rules we have tested.

  • Search application, proxy, and WAF logs for requests touching /or.

References

1 on the record

Elsewhere on this site

  • baxterevery CVE for this vendor
  • CWE-306other pages naming this weakness

Not in any source we poll

Listed rather than left blank: an empty field and an unmeasured one look identical on screen, and only one is a reason to look elsewhere.

  • No confirmed IOCs, IP addresses, domains, file hashes, or malware artifacts supplied.
  • No organization-specific asset inventory, compensating-control status, or patch deployment evidence supplied.
  • No exploit packet captures, log samples, or incident case IDs supplied.